![]() Solution:
1. Check whether the communication line between the computer and the testing machine is off;
2. Check whether the online selection sensor is selected correctly;
3. Check whether the sensor has been hit during the recent test or operating the keyboard;
4. Check whether the software has been used before the problem occurs. Calibration or calibration function;
5. Check whether the calibration value, calibration value or other information in hardware parameters has been manually changed.
Question 2: The power supply of the main body of the spring tension and compression testing machine is not bright, and it cannot move up and down. Solution:
1. Check whether the power line connected to the testing machine is connected normally;
2. Check whether the emergency stop switch is in the screwed state;
3. Check whether the power supply voltage connected to the testing machine is normal;
4. Check whether the fuse on the machine socket is blown , please take out the spare fuse and install it.
Question 3: The main body of the spring tension and compression testing machine is powered but the device cannot move up and down.Solution:
1. Check whether the device can not move after 15S, because the host needs to self-check when it is turned on, which takes about 15S;
2. Check whether the upper and lower limit positions are in proper positions, and there is a certain operating space;
3. Check the access to the testing machine Whether the power supply voltage is normal.